Measuring range:
Carbon: 0.00001-10.0000% (it can be expanded to 99.999%)
Sulfur: 0.00001-10.0000% (it can be expanded to 99.999%)
Analysis time: 25-60s adjustable; it is usually about 35s (excluding sampling and weighing time)
Measuring pool: carbon measuring pool, sulfur measuring pool. (For carbon measurement it can be made into double carbon pools according to the users’ needs, with high and low carbon pools switched automatically.)
Analysis error, accuracy and precision:
Analysis error: conforms to GB/T20123-2006, ISO15350:2000
Precision: conforms to the national metrological verification regulation JJG395-97
Analytical accuracy: carbon: 1ppm, sulfur: 1ppm
Sensitivity (min reading): 0.1ppm
High frequency induction furnace combustion power: 2.5-3.5KVA adjustable. Oscillation frequency: 20MHz
Electronic balance: on-line operation, non-quantitative weighing.
Computer: mainstream model and 19-inch LCD
Working power supply: AC220V ±10% 50Hz
Working environment: indoor temperature: 10-30 ℃, relative humidity: ≤ 80%
Oxygen: purity ≥ 99.5%, input pressure 0.18MPa, oxygen carrier pressure 0.08MPa
Gas flow rate: top oxygen flow rate 1. 0-2.0L 4.5L/min; Analytical gas flow rate 3.0-4.5L/min.
Desiccant: highly efficient discoloration desiccant
Filter agent: quartz cotton