1. Power supply: AC220V ±10%, 50Hz
2. Output voltage:0~60kV, 80kV, 100kV
3. Capacity: 1.6kVA, 2.0kVA
4. Voltage rise rate: adjustable 0.5kV/s, 1kV/s, 2 kV/s, 3kV/s, 5kV/s.
5. Testmethods: “IEC156/IS6792 etc.” “Custom Test”, “ASTM D877”, “ASTM D1816”, “Manual”, “5 minutes test”, “Proof A”, “Proof B” and “UNE21”.
(Proof A: Voltage rises at 2kV/s to set value, stays for 60s unless breakdown occurs earlier and voltage falls.
Proof B: As above and continues to breakdown or maximum value of test set.)
6. Voltage inspection accuracy: <2%
7. Breakdown sensitivity: <0.1kV
8. Waveform distortion: ≤1%
9. Breakdown response time: ≤10ms