1 |
Measurement range |
(0~600) penetration scale |
2 |
Timing controller |
5s,8s,10s,12s,30 sand 60s,bias within±0.1s |
3 |
Resolution |
0.01mm |
4 |
Penetration accuracy |
±1penetration |
5 |
Heater power |
200W |
6 |
Temperature control accuracy |
25±0.1℃ (Note:The ambient temperature shall not exceed 20 ℃) |
7 |
Temperature control mode |
High accuracy digital temperature controller |
8 |
Constant temperature bath |
Hard glass bath |
9 |
Stirring |
automatically stirred by a magnetic stirrer |
10 |
Standard needle |
2.5±0.05g |
11 |
Standard cone |
102.5±0.05g |
12 |
Needle holder |
Adjustable. Convenient toad just and exact to aim the sample |
13 |
Others |
Equipped with cold light source and magnifier; Convenient for operation and observation |